YSI 5200A
123
YSI 5200A
122
Display Settings
Menus → System → Display Settings
Sensor and Aux conditions are displayed at the Run Screen. Use the Display Set-
tings menu to configure how many systems display on one page at the Run Screen.
Up to six systems can be displayed on one Run Screen page. Multiple Run Screen
pages scroll at user configured speed (1-65535 seconds).
1. Enter Display Settings menu
Menus → System → Display Settings →Display Forma
t
2. Use
▲
and
▼
to scroll and highlight submenus.
Configuring the 5200A
3. Press
to select.
4. Press
to save configuration and return to Run Screen.
Notes:
-
At 1x2 & 2x2 display settings, analog aux values over 7 digits will overlap on to
the adjoining sensor window. Adjust aux display format to lower resolution.
-
Because of font limitations, 1X2 and 2X3 display formats do not show all 20
characters of a 20 character name. Typically only 13 characters are displayed.
Updating 5200A Firmware
Like all microprocessor based instruments, it may be necessary to upload revisions
to the 5200A firmware. Download new firmware using the Firmware Updater utility
in AquaManager. A serial RS232 connection is necessary to run the Downloader
program.
The instrument firmware may become corrupted if any of the following occurs
during a 5200A firmware update:
-
Firmware Updater utility is closed;
-
AquaManager is closed;
-
RS232 cable is removed from 5200A or PC com port;
-
power is disconnected from 5200A.
If the 5200A’s firmware becomes corrupted and is not updated properly, the instru-
ment will need to be returned to YSI
.
Firmware Updater Directions
1. Wire RS232 connection. See RS232 Communication - page 58. (Dis-
regard if already using RS232 connection.)
Note:
Connect RS232 directly into slaves in order to upload firmware to slaves.
Disable slaves at the master to avoid slave failure alarms when uploading
-
Menus → System → Communications → Network →Master.
2. Enable the Force Serial Port on 5200As wired and configured with an
ethernet device - page 101.
3. Verify that the baud rate in 5200A Communications menu is configured
to 115200.
4. From the AquaManager Explorer window, access the Firmware Updater
utility from the pull down menu: Tools→ Instrument tools→Update
Firmware.
Configuring the 5200A