Notes:
-
Logs cannot be deleted from the 5200A instrument using AquaManager.
-
Calibration and Configuration
logs cannot be cleared from 5200A.
-
Clear data logs are updated immediately to the system and not when
is pressed. YSI recommends pressing
in this instance in order to save
any other changes that may have been made to the configuration.
Date/Time
Menus → System → Date/Time
It is important to configure the correct date and time in the 5200A. Four 10-event
daily timers and one feed timer run based on the 5200A date and time. Logs record
events based on the 5200A clock. Date and time display formats are user defined.
1. Enter Date/Time menu
Menus → System → Date/Time
2. Verify 5200A current date and time.
3. If necessary, use
▲
and
▼
to scroll and highlight submenus.
4. Press
to select.
5. Configure submenus.
6. Press
to save setting and return to Run Screen or press
to System menu.
Notes:
-
Changes to Date and Time menus take place immediately and are saved prior
to
.
-
Configured date and time formats are carried throughout menus and logs.
-
Any active (with energized relay) timer system, including 10-event daily
timers, feed timers, and/or dosing timers, are NOT recalculated when the
5200A time and/or date is changed. Once a timer relay has been energized, it
runs for the duration and does not reset based on changes to the 5200A date
and time. Example - timer 1, event 1 is configured to come on at 9:00am for
Configuring the 5200A
10 minutes. Timer relay energizes at 9:00am. At 9:05am the 5200A time is
changed to 10:05am (for daylight savings adjustment). The timer relay will
de-energize after five additional minutes. The 5200A time will be 10:10am.
The event log will reflect the configured 5200A time when the event occurred.
In this example, 9:00am for the start and 10:10am for the end.
-
The time must be manually changed for daylight savings time adjustments.
Daylight savings note: Sensor data log records are not overwritten when the
clock is changed unless the instrument is power cycled after changing the
data and time. Example - the data log interval is set to 15 minutes. Current
instrument time is 6:00 am. The clock is adjusted to 5:00 am for daylight
savings time. The 5200A sensor log contains four 15 minute interval records
from 5:00 am to 6:00 am from prior to time change. Sensor data logs are not
overwritten with new values for the repeat of 5:00 am to 6:00 am 15 minute
intervals unless the instrument is power cycled.
-
Menu configurations are saved to the system by pressing
to return to
the Run Screen and not when the change is made in the menu. Exception -
Date and time and clear data logs are updated immediately to the system, not
when
is pressed.
Configuring the 5200A