YSI 5200A
125
YSI 5200A
124
5. Configure PC com port number (1-8) in Flash Programmer.
Notes:
-
Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI
p/n
773
.
-
When using USB-Serial Adapter, verify that the port number assigned dur-
ing driver installation is the same port configured when using AquaManager
Flash Programmer.
6. AquaManager Program Folder includes the latest version of instru-
ment firmware. The lastest firmware is stored as an .mdb file in the PC
Program Files>AquaManager>Firmware. At AquaManager, Firmware
Updater>File>Load Firmware Database. Locate and load this file into the
Firmware Updater.
7. Left click
button.
Configuring the 5200A
window opens.
8. Select “Yes” and the following screens will appear.
On
instrument
On
PC
YSI 5200A/5400/5500 Bootstrap Loader screen
Flash Programmer uploading firmware into
5200A
Note:
-
If the YSI 5200A/5400 Bootstrap Loader screen does not appear at the in-
strument after above steps manually place 5200A in downloader at
Menus
→ System → Downloader.
Close and reopen AquaManager. Reopen Flash
Programmer and repeat steps starting at 6.
-
Power cycle to exit downloader and return to run screen. Do not power
cycle during an active upload.
-
If after manually placing the 5200A into downloader and the Flash Program-
mer utility still does not upload firmware see Force Downloader - page 127
.
Configuring the 5200A