YSI 5200A
103
YSI 5200A
102
Probe Name
Menus → System → Configure Aux →Aux 1-2→Probe Name
Use the alpha/numeric keypad to name aux inputs. Configured name(s) appear
throughout menus and at Run Screen - figure 4.6, page 106.
Note:
-
Aux names can also be configured at Menus → System → Name Devices/Sensors.
Input Type
Menus → System → Configure Aux →Aux 1-2→Input Type
Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup (page
157) menu selections are based on the aux input type configured at the System menu.
Select the analog input type based on the signal being sent from the device.
Notes:
-
It is normal to hear an on-board relay “click” when switching to or from the
4-20mA aux input type.
-
Analog systems will not monitor, control, or alarm properly if a device that
operates outside the configured analog range is installed and configured.
Configuring the 5200A
Digital
Menus → System → Configure Aux →Aux 1-2→Input Type→Digital
Use the digital input when monitoring an external device to control or alarm when
the contacts of the external switch are open or close. A common application would
be to detect high or low water levels using a float switch.
Notes:
-
Suffix, Min, Max, Input Scale Min, Input Scale Max and Display Format are
not available for digital inputs.
-
In the 5200A sensor log, aux digital “closed” is logged as Aux 0.00. Aux digital
“open” is logged as Aux 1.00.
1. Wire aux digital device - page 49.
2. Enter Aux menu
Menus → System → Configure Aux →Aux 1-2
3. Use
▲
and
▼
to scroll and highlight submenus.
4. Press
to select.
5. Configure digital input type and name.
6. Press
to save configuration and return to Run Screen.
7. Go to Sensor Setup Menus → Sensor Setup Aux 1-2 - page 162 to enable
and configure control/alarm on condition, control/alarm relay, and hold
off time.
8. Press
to save configuration.