Email Retry Delay
Menus → System → Communications → Email Retry Delay
Configure the Email Retry Delay for the time interval that emails should be retrans-
mitted after the initial alarm email is sent. The initial alarm email is sent immediately
after the general alarm hold off time or after the aux hold off time. Emails will be
retransmitted until one of the bottom front panel softkeys is pressed, regardless if
the alarm condition still exists.
Note:
-
On a network, emails are retransmitted according to the slave retry delay
configuration of the slave in alarm and not the master’s email retry delay
configuration. Please note, a slave Email Retry Delay menu will display (Not
available) because of its slave configuration; however, this feature is available
and has a default value of 15 minutes that can be configured by the user.
-
The Retry Delay should be configured at intervals ≥ 15 minutes.
-
AquaManager alarm emails are retransmitted based on the 5200A slave(s) and
master email retry delay configuration.
Baud Rate RS232
Menus → System → Communications → Baud Rate
Configure the Baud rate (for RS232 connections only) in the 5200A to match the
baud rate configured in AquaManager. Use the default rate of 115200 for fastest
communication speed.
1. Enter Network menu
Menus → System → Communications → Network →Baud Rate
2. Press
▲
or
▼
to scroll and highlight a selection.
3. Press
to confirm selection.
4. Press
to save configuration and return to Run Screen.
Configuring the 5200A
Force Serial Port
Menus → System → Communications → Force Serial Port
Enable the Force Serial Port when the 5200A is communicating via RS232 (direct)
to a PC and when the 5200A has a serial to ethernet device.
1. Wire RS232 connection - page 58
2. Enter Network menu
Menus → System → Communications → Force Serial Port
3. Use the
softkeys to configure off or on.
4. Press
to save configuration and return to Run Screen.
Notes:
-
RS232 connection is required to upload firmware using the Firmware Updater.
-
The RS232 icon
blinks at the Run Screen when in Force Serial Port mode.
Non-Networked (Stand Alone) 5200As
Non-networked (stand alone) 5200As must have the termination relay enabled. In
addition, non-networked (stand alone) 5200As must be configured as a master in
the System Communication menu with no slave nodes enabled.
Configure AUX
Menus → System → Configure Aux 1-2
The 5200A has two auxiliary (Aux) inputs. Aux inputs can be configured as digital or
analog (0-1VDC, 0-5VDC, or 4-20 mA) inputs. Configure suffix, minimum (Min),
maximum (Max), Input Scale Min, Input Scale Max and display format menus for
analog inputs. Go to Sensor Setup Menus → Sensor Setup Aux 1-2 - page 161 to
enable Aux system and configure control/alarm settings. Both digital and analog
directions follow.