YSI 5200A
59
YSI 5200A
58
Step 8 Wire Communication Equipment
Communication equipment must be installed in order to communicate, either lo-
cally or remotely, with the 5200A. The communication connection can be made
locally (RS232) or remotely with a serial to ethernet device.
Only one connection
method can be used at a time.
AquaManager supports both connection methods.
Notes:
-
Firmware can only be uploaded to an instrument using a local RS232 con-
nection. See Update 5200A Firmware - page 123.
-
On a network, install and configure RS232 or the serial to ethernet device
hardware on the master node only.
Email Alarming
Email alarming can be configured in one of two ways:
-
from a PC running the AutoPoll Program that is included with AquaMan-
ager software (PC must have internet connection). This is referred to as
AquaManager email
.
-
via a serial to ethernet* device with TCP/IP connection at master 5200A.
This is referred to as
5200A email
.
*Serial to ethernet connection requires access to a SMTP server accessible by the
local network.
RS232 Communication
Use Category 5e patch cable from RS232 port at location
H
- figure 3.29 of I/O
board to PC using RJ45 to DB9 adapter.
Directions to Wire D
irect RS232 PC
Connection
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer
un câblage quelconque.
1. Perform steps 1 - 4 of 5200A installation and steps 5, 6 and 7 as required
for the application.
2. Open front panel - page 40.
Installation and Wiring
3. Feed direct connect cable through drilled hole of rubber grommet and
to location
H.
4. Run high and low voltage cables through separate bulkhead and conduit.
5. Connect one end of communication cable to RJ45 at location
H.
Connect
other end to PC com port using RJ45 to DB9 adapter.
B
C
D
E
F
G
H
K
L
D
o not c
onnec
t or disc
onnec
t IDC r
ibbon cable
when 5200A is po
w
er
ed
. S
er
ious damage can oc
cur
.
Figure 3.29
6. Complete other wiring to 5200A.
7. Close front panel - page 40.
8. Test RS232 communication.
Notes:
-
When using a USB-Serial Adapter, verify that the port number assigned
during driver installation is the same port configured when using Aqua-
Manager and/or the Firmware Updater.
-
Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI
p/n
773
(64-bit).
-
When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232
Line Extender/Booster. Mount the booster in an area where it is protected
from moisture.
Installation and Wiring