YSI 5200A
107
YSI 5200A
106
Display Format
Menus → System → Configure Aux →Aux 1-2→Display Format
Configure the Display format (0, 0.0, or 0.00) based on the resolution to be displayed
on the Run Screen.
Example - when using a total dissolved gas pressure sensor configure, the display
format as 0.0 - figure 4.6, page 106.
In the below screen shot, the Run Screen is displaying aux name “TGP,” suffix
“mmHg,” with display format “0.0”. The display value represents an input of 1.875
volts with 0-5VDC analog aux configuration. The min value = “-400”. The max value
= “+400”. The control/alarm low value is “-50”. Alarms are enabled for this aux input.
figure 4.6
Configuring the 5200A
Notes:
-
Values are displayed based on the display format resolution. Example - if the
display format is 0.0 and the min value configured is 1.88, then 1.9 will be the
value displayed in the min value menu.
-
5200A aux sensor logs record 0.00 resolution regardless of the selected resolu-
tion in the aux analog display format configuration.
Directions to configure Aux analog - System>Configure Aux menu
1. Wire aux analog devices - page 49.
2. Enter Aux menu
Menus → System → Configure Aux →Aux 1-2
3. Use
▲
and
▼
to scroll and highlight submenus.
4. Press
to select.
5. Press
to save configuration.
6. Go to Sensor Setup Menus → Sensor Setup Aux 1-2 - page 162 to enable
and configure set points, control, alarm, relays, and hold off time.
7. Press
to save configuration.
Configure E-Aux
Menus → System →Configure E-Aux 1-8
5200A instruments have eight E-Aux systems. E-Aux systems are sometimes
referred to as virtual system because unlike Aux systems which use two onboard
(Aux) inputs for wiring input devices, E-Aux systems use inputs that are wired to a
networked IOEM. 5X00 E-Aux systems mimic aux systems but use a remote IOEM
wiring location for the actual input. E-Aux inputs 1-8 can be configured for digital,
analog ( 0-1V,DC 0-5VDC, or 4-20 mA) or temperature inputs. System → Configure
E-Aux 1-8 menu selections change based on input type selection. Configure suffix,
minimum (Min), maximum (Max), Input Scale Min, Input Scale Max and display
format menus for non temperature analog inputs. To configure setpoints,control,
alarms and relays see specific sensor setup menus for configured input type. For E-
AuxAnalog Sensor Setup - page 163 . For E-Aux Temperature Sensor Setup - page
165. For E-Aux Digital Sensor Setup - page 162.
NEW
Configuring the 5200A