2-4
IM WT5000-01EN
Comment (Comment)
You can add a comment that consists of up to 30 characters when you save a file. You do not have to enter a
comment. You can use all characters and spaces that are displayed on the keyboard.
Settings That Are Shared with Those of Other Menus
The following settings are shared by the menus for saving setup data, storing data, saving data.
• File list display and save destination settings (File List)
• Automatic file naming settings (Auto Naming)
• File name (File Name)
The comment settings (Comment) are shared by the menus for saving setup data and storing data.
Saving (Save Exec)
Saves the data to the specified save destination with the specified file name.
Loading Setup Data (Load Setup, LOAD(SETUP))
On the file list, specify a file to load setup data from. The extension is .SET. For information about how to
configure the file list display and how to operate files and folders, see “File List (File List).”
Loading (Execute Load)
Loads the data of the specified file.
• If you change the extension of the saved data file, by using a PC or some other device, the instrument will
no longer be able to load it.
• When setup data is loaded from a file, the setup data for each key is changed to match the loaded settings,
and it cannot be changed back. We recommend that you save the current settings before loading different
setup data.
• The date, time, and communication setup parameters are not saved. So even if you load setup data from a
file, the date, time, and communication settings will not change.
• This instrument cannot load setup data that has been saved by a product with an incompatible firmware
version.
• This instrument cannot load setup data that was saved by an instrument with a different element
configuration, different options, etc.