19-8
IM WT5000-01EN
GP
−
IB
Connects this Instrument to a PC using GP-IB.
Address (Address)
• You can select a value between 0 and 30.
• Each device that is connected by GP-IB has its own unique address in the GP-IB system. This address is
used to distinguish between different devices. Therefore, you must assign a unique address to the instrument
when connecting it to a PC or other device.
When the controller is communicating with the instrument or with other devices through GP-IB, do not change
the address.
Notes about Connections
• Multiple cables can be used to connect multiple devices. However, no more than 15 devices, including the
controller, can be connected on a single bus.
• When connecting multiple devices, you must assign a unique address to each device.
• Use cables that are no longer than 2 m in length to connect devices.
• Make sure the total length of all cables does not exceed 20 m.
• When devices are communicating, have at least two-thirds of the devices on the bus turned on.
• To connect multiple devices, use a star or daisy-chain configuration. Loop and parallel configurations are
not allowed.
USB(USB
−
TMC)
Connects this Instrument to a PC using USB.
To connect the instrument to a PC through the USB port, carry out the following procedure.
Install the YOKOGAWA USB driver on your PC. For information about how to obtain the YOKOGAWA USB
driver, contact your nearest YOKOGAWA dealer. You can also access the YOKOGAWA USB driver download
webpage and download the driver (https://tmi.yokogawa.com/library/).
WT-Compatible Command Type (Command Type)
Many of the functions of this instrument can be controlled using the communication commands of other WT
series instruments.
For details, see the Communication Interface User's Manual, IM WT5000-17EN.
Releasing Remote Mode (UTILITY)
Press UTILITY to switch from remote mode (in which the REMOTE indicator is illuminated) to local mode (in
which front panel key operations are valid). This key is disabled when the instrument is in local lockout mode.
19 Utility