17-4
IM WT5000-01EN
Conditions for Saving Numeric Data
You can save the numeric data that this instrument has measured to a file in ASCII format (.csv).
Saved Numeric Items (Saved Numeric Items)
Select the type of items to save from the following:
• Displayed numeric items (Displayed Numeric Items)
• Selected items (Selected Items)
Displayed numeric items (Displayed Numeric Items)
The saved items vary as indicated below depending on the display.
• When numeric values are displayed in the 4-, 8-, or 16-value display or the matrix display
All the measurement functions on the page that is displayed when saving starts are saved in the order that
they are displayed.
• When numeric values are displayed in the harmonics list single screen or harmonics list dual screen
In addition to the data described above, the data of harmonics that are not displayed on the screen is saved
up to the maximum measurable order (Max Order).
• When numeric values are displayed in the all display
The data cannot be saved. If you attempt to save the data, an error message is displayed.
• In non-numeric displays (e.g., waveform display, trend display)
Numeric data is not saved.
Selected Items (Selected Items)
You can select the types of numeric data to save.
Selecting Numeric Items to Save (Select Saved Numeric Items)
If you selected Selected Items, select which numeric data to save.
All ON (All ON)
The data of all measurement functions is saved.
All OFF (All OFF)
None of the data for any of the measurement functions is saved.
Preset 1 (Preset 1)
The data of the following measurement functions is saved for all elements and wiring units.*
Urms, Irms, FreqU, FreqI, P, S, Q, λ, Φ
Preset 2 (Preset 2)
The data of the following measurement functions is saved for all elements and wiring units.*
WP, WP+, WP−, q, q+, q−, ITime, WS, WQ
Element (Element/Σ)
To select an element or wiring unit that you want to save the data of, select or clear its check box. You can select
from the following options.
Element1, Element2, Element3, Element4, Element5, Element6, Element7, ΣA, ΣB, ΣC
* If the wiring system setting (Wiring) is configured so that a wiring unit does not exist, the data for the functions
of that wiring unit is not saved. For example, if ΣC does not exist, the data for ΣC is not saved.
Motor Number (Motor)
Select whether to save the data of motor input 1 to 4 by selecting or clearing their check boxes.
Auxiliary Input (Aux)
Select whether to save the data of auxiliary input 1 to 8 by selecting or clearing their check boxes.
17 Saving Numeric Data, Waveform Data, and Screen Images