Real-Time Normal Integration Mode (R–Normal)
Set the date and time when integration starts and stops and the duration of integration. Integration starts at the
scheduled start time. When one of the conditions below is met, integration is stopped, and the integration time
and integrated value are held.
• The scheduled stop time is reached.
• The specified timer time elapses.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
Integration time
Integrated value
Hold
Hold
Start
Date and time
Stop
Date and time
Reset
In real-time integration mode, when you set the integration time to 00000:00:00, integration starts at the
scheduled start time. When one of the conditions below is met, integration is stopped, and the integration time
and integrated value are held.
• The scheduled stop time is reached.
• The Stop button is tapped, or the STOP key is pressed.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
14 Integrated Power (Watt Hours/Ampere Hours)