5-7
IM WT5000-01EN
Analog Input Range
Set the analog input range for input signals whose type has been set to analog. For input signals whose type
has been set to pulse, there is no need to set the analog input range.
Analog Input Auto Range (Analog Auto Range)
Select whether to turn auto range on or off. When auto range is on, this instrument automatically switches
between the following ranges according to the size of the input signal.
20 V, 10 V, 5 V, 2 V, 1 V
Range Increase
• The measurement range is increased when the input signal exceeds 110% of the measurement range.
• The measurement range is increased when the peak value of the input signal exceeds approximately150% of
the measurement range.
Range Decrease
The measurement range is decreased when the input signal falls below 30% of the measurement range and the
peak value of the input signal is less than 125% of the lower range.
When non-periodic pulse waveforms are applied during auto range, the range may not remain constant. If this
happens, use the fixed range setting.
Analog Input Range (Fixed Range) (Analog Range)
You can select one of the following input ranges.
20 V, 10 V, 5 V, 2 V, 1 V
5 Motor Evaluation and Auxiliary Input (Option)