14-2
IM WT5000-01EN
At Integration Ready
In real-time integration mode, after you tap Start or press START key, if the time is before the scheduled
integration start time, and the scheduled integration start time are displayed.
Scheduled integration start time
At Integration Timeout
If the integration timer expires, integration is stopped automatically, and and the elapsed integration time are
displayed.
Error
In the following situations, the integration will automatically stop and and the elapsed integration time will
appear.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
• A power failure occurs when integration is in progress. Even when a power failure occurs, this instrument
stores and holds the integration result. Set the
integration resume action at power failure recovery
to specify
the action that will be taken when the power recovers after a power failure.
Example of the Display When the Integration Time Is Set to a Non-Zero Value
Elapsed integration time
Integration timer
Example of the Display When Independent Integration Is Enabled
Integration status of each element
Element 1: Running
Element 2: Running
Element 3: Reset
Element 4: Stop
Element 5: Running
Element 6: Stop
Element 7: Stop
14 Integrated Power (Watt Hours/Ampere Hours)