16-13
IM WT5000-01EN
When the Storage Mode Is Integration Synchronization and the Integration Mode Is Continuous or Real-
Time Continuous
Storage is performed at the first data update after the storage interval passes. Storage continues even when
integration is reset by the integration timer. When the integration timer is reset, the storage interval timer is also
reset. The figure below is an example for when the integration timer is set to 20 s.
Time [s]
Data
update
Storage
Time [s] 0 5 10 15 20 5 10 15
0 7 14 20 7 14
• When the storage of numeric data at store start (Store At Start) is enabled, storage starts when the time is
at 0 s (store start).
• Storage is performed when the integration mode is continuous integration mode or real-time continuous
integration mode, the integration timer time elapses, and the integrated values are reset. In the above
example, the integration timer is set to 20 s, so the data is stored after 20 s have passed, and then the
integrated values are reset.
16 Data Storage