Stored Items
You can set which numeric data items to store.
Stored Items (Stored Items)
This is fixed to selected items (Selected Items).
Selected Items (Selected Items)
You can select the types of numeric data to store. To select which types of numeric data to store, use items (Select
Stored Items).
Select Stored Items (Select Stored Items)
If you set the numeric data items to store to Selected Items, you can select which types of numeric data to store.
All ON (All ON)
The data of all measurement functions is stored.
All OFF (All OFF)
None of the data for any of the measurement functions is stored.
Preset 1 (Preset 1)
The data of the following measurement functions is stored for all elements and wiring units.*
Urms, Irms, FreqU, FreqI, P, S, Q, λ, Φ
Preset 2 (Preset 2)
The data of the following measurement functions is stored for all elements and wiring units.*
WP, WP+, WP−, q, q+, q−, ITime, WS, WQ
Element (Element/Σ)
To select an element or wiring unit that you want to store the data of, select or clear its check box. You can select
from the following options.
Element1, Element2, Element3, Element4, Element5, Element6, Element7, ΣA, ΣB, ΣC
* If the wiring system setting (Wiring) is configured so that a wiring unit does not exist, the data for the functions
of that wiring unit is not stored. For example, if ΣC does not exist, the data for ΣC is not stored.
Motor Number (Motor)
Select whether to store the data of motor input 1 to 4 by selecting or clearing their check boxes.
Auxiliary Input (Aux)
Select whether to store the data of auxiliary input 1 to 8 by selecting or clearing their check boxes.
16 Data Storage