14-8
IM WT5000-01EN
Real-Time Continuous Integration Mode (Continuous integration; R-Continuous)
Set the date and time when integration starts and stops and the duration of integration. Integration starts at the
scheduled start time. When the specified timer time elapses, integration is automatically reset and restarted. If
either of the conditions below is met, integration is stopped, and the integration time and integrated value are
held.
• The scheduled stop time is reached.
• The Stop button is tapped, or the STOP key is pressed.
• The integrated value reaches its maximum or minimum displayable value.
Integration time
Integrated value
Hold
Hold
Timer
value
Timer
value
Timer
value
Start
Date and time
Stop
Date and time
Reset
Integration Timer (Integration Timer)
You can set the hour, minute, and second within the following range.
00000: 00: 00 to 10000: 00: 00
When Independent Integration Is Disabled
The integration timer that you specified above applies to every element.
When Independent Integration Is Enabled
• Setting (Setting)
You can select the method for setting the integration timer from the following options.
• Each (Each): Set it separately for each input element.
• All (All): Set it for all installed input elements at the same time.
• Element1 to Element7
You can set the integration timers for each element within the range given above.
14 Integrated Power (Watt Hours/Ampere Hours)