16-9
IM 253421-01E
16
Specifications
16.14 General Specifications
Item
Specification
Warm-up time
Approx. 30 min.
Ambient temperature
5 to 40
°
C, 20% to 80% R.H. (no condensation)
and humidity range
Operating altitude
2000m or below
Storage temperature
–25 to 60
°
C (no condensation)
Insulation resistance
Between voltage input terminals and case
Between current input terminals and output terminals
Between voltage input terminals and current input terminals
Between voltage input terminals and power plug
Between current input terminals and power plug
Between case and power plug
Above: 50 M
Ω
or more at 500 V DC
Withstanding voltage
Between voltage input terminals and case
Between current input terminals and output terminals
Between voltage input terminals and current input terminals
Between voltage input terminals and power plug
Between current input terminals and power plug
Above: AC 3700 V for 1 minute at 50/60 Hz
Between case and power plug: AC 1500 V for 1 minute at 50/60 Hz
Power supply
100 to 240 V; frequency: 50/60 Hz
Vibration test condition
Sweep test - Frequency:
8 to 150 Hz sweep, all 3 directions for 1 minute
Endurance test - Frequency:
16.7 Hz, all 3 directions; amplitude of 4 mm for 2 h
Impact condition
Impact test: Acceleration at 490 m/s
2
, all 3 directions
Free-fall test - Height: 100 mm, 1 time for each 4 sides
Power consumption
25 VA maximum (Power supply: 120 VAC)
35 VA maximum (Power supply: 240 VAC)
External dimensions
Approx. W
×
H
×
D : 213
×
88
×
350 (mm), 8-3/8
×
3-1/2
×
13-3/4 (inch), excluding projections.
Weight
Approx. 3.0 (kg), 6.6 (lbs)
Accessories
Power cord: UL/CSA, VDE, SAA or BS standard 1 pc
24-pin connector
User’s Manual
Rubber feed
Emission
Complying Standard: EN55011-Group1, Class A
This is a Class A product for industrial environment. In a domestic environment, this product
may cause radio interference in which cause the user may be required to take adequate
measures.
Cable Condition:
External Senser Input (installed /EX1 or /EX2 option)
500 mm max
External Input/Output Signals (installed /DA4, /CMP option)
To use shielded wires
Immunity
Complying Standard: EN50082-2:1995
Safety standard
Complying Standard: EN61010
Overvoltage Category II
Pollution degree 2
*
Equipment that is connected to the WT200 through the GP-IB, RS-232-C, or Ext. I/O connector must comply with applicable
safety standards (IEC60950 or IEC61010-1, for example) or some equivalent standard.