7-5
IM 253421-01E
7
Using the Harmonic Analysis Function (Optional)
7.3
Switching the Harmonic Analysis Function ON/OFF
Keys
DC
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
M W
h
k
A
h
m V
Hz
M W
k
A
deg
var
TIME
FUNCTION
AUTO
MODE
TRIG
MAX HOLD
HOLD
CAL
INTEGRATOR
HARMONICS
INTERFACE
OUTPUT
SHIFT
REMOTE
MEMORY
INTEG SET
ENTER
A RANGE
V RANGE
START
LOCAL
SETUP
STOP
RESET
AUTO
FUNCTION
FUNCTION
m V
PF
M W
k
A
m V
VA
V MEAN
RMS
A OVER
V OVER
SAMPLE
A
B
C
MODE
Displays
relevant
keys and
indicator
Procedure
•
Operate the instrument by following the thick lines in the menu below.
•
Press the ENTER key to confirm a selection or setting.
•
To leave the current menu in the middle of the operation, press the key indicated in step 1.
The confirmed settings up to that point are kept.
1.
SHIFT
START
HARMONICS
2.
ENTER
3.
• Turning the Harmonic Analysis Function ON/OFF
End
(Display C)
Explanation
Turning the Harmonic Analysis Function ON/OFF
• on : Pressing the ENTER key after selecting on will result in starting of the harmonic
analysis and the HARMONICS indicator will light up. The harmonic order will be
displayed on display A.
• oFF: Pressing the ENTER key after selecing off will result in stopping of the harmonic
analysis and the HARMONICS indicator will extinguish.
Note
•
When the harmonic analysis function is turned ON, the measurement mode will automatically
change to RMS mode. When the harmonic analysis function is turned OFF, the measurement
mode will stay the RMS mode.
•
When the harmonic analysis function is ON, integration cannot be started. And accordingly,
when the integration is in progress, the harmonic analysis function cannot be started (see
page 6-10).