13-11
IM 253421-01E
Communication Commands1 (System of Commands before the 488.2-1987 Standard)
13
SN/SN?
Sets the measurement synchronization
source or queries the current setting.
Syntax
SN m<terminator>
“m” indicates the type of measurement
synchronization source.
m=
0: V (voltage)
1: A (current)
Query
SN?<terminator>
Example
SN0
Description
Parameter error 12 will occur if “m” is set to an
illegal value.
SO/SO?
Sets the store function ON/OFF or
inquires about the current setting.
Syntax
SO m<terminator>
“m” indicates whether storage is ON or
OFF.
m=
0:OFF
1:ON
Query
SO?<terminator>
Example
SO1
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SR/SR?
Sets the storage interval or queries the
current setting.
Syntax
SR m1,m2,m3<terminator>
“m1” indicates the hour
0
≤
m1
≤
99
“m2” indicates the minutes
0
≤
m2
≤
59
“m3” indicates the seconds
0
≤
m3
≤
59
Query
SR?<terminator>
Example
SR0,0,0
Description • Parameter error 12 will occur if an illegal value
is set.
• When the storage interval is set to 0hrs, 0min,
0sec, see page 8-2.
• While recalling or storing is in progress,
execution error 19 will occur.
SS
Stores setting parameters into a
selected file.
Syntax
SS m<terminator>
“m” indicates file no., and must be set
within the following range.
1
≤
m
≤
4
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
• The following setting parameters can be stored:
All setting parameters which can be output by
the OS command
Information related to communications (GP-
IB, RS-232-C etc.)
• While recalling or storing is in progress,
execution error 19 will occur.
TM/TM?
Sets integration preset time or queries
the current setting.
Syntax
TM m1,m2,m3<terminator>
“m1” indicates hour, and must be set
within the following range.
0
≤
m1
≤
10000
“m2” indicates minute, and must be set
within the following range.
0
≤
m2
≤
59
“m3” indicates second, and must be set
within the following range.
0
≤
m3
≤
59
Query
TM?<terminator>
Example
TM0,0,0
Description • Parameter error 12 will occur if an illegal value
is set.
• While recalling or storing is in progress,
execution error 19 will occur.
• The maximum selectable time is 10000 (hours).
WR/WR?
Sets the wiring system or queries the
current setting.
Syntax
WR m<terminator>
m=
1:
1
Φ
2W
Query
WR?<terminator>
Example
WR1
Description
Parameter error 12 will occur if an illegal value
is set.
YC/YC?
Sets the display channel while the
comparator function is ON or queries
the current setting.
Syntax
YC m<terminator>
“m” indicates the channel number for
display
• During single mode
m=
1:
Displays limit and measurement
value on display 1
2:
Displays limit and measurement
value on display 2
3:
Displays limit and measurement
value on display 3
4:
Displays limit and measurement
value on display 4
• During dual mode
m=
1,2:Displays limit and measurement
value on display 1 and 2
respectively
m=
3,4:Displays limit and measurement
value on display 3 and 4
respectively
Query
YC?<terminator>
Example
YC1
Description
Parameter error 12 will occur if “m” is set to an
illegal value.
YM/YM?
Sets the mode of the comparator
function or queries the current setting.
Syntax
YM m<terminator>
“m” indicates the display mode
m=
0:
Single mode
1:
Dual mode
Query
YM?<terminator>
Example
YM0
13.1 Commands