15-4
IM 253421-01E
Communication Commands to Carry Out Adjustments
Command
Description
CAL1
Enters the range adjustment mode.
CAL2
Enters the external input range adjustment mode.
CAL3
Enters the D/A output adjustment mode.
CAL0
Finishes adjustment (and returns to normal measurement mode).
CR0
Switches to 30 V range in range adjustment mode.
CR1
Switches to 300 V range in range adjustment mode.
CR2
Switches to 1 A range in range adjustment mode.
CR3
Switches to 10 A range in range adjustment mode.
CR6
Switches to 100 mA range in range adjustment mode.
CHm
Switches the channel in D/A output adjustment mode.
m = 1 to 4
CDm,n
Enters the actual output value in D/A output adjustment mode.
m = 1 to 4, n = actual output value
DO0
O5 V in D/A output adjustment mode.
DO1
Outputs –5 V in D/A output adjustment mode.
OD
Requests the output of measurement data, and sets the output format to normal
measurement default.
ENT
Corresponds to the ENTER key operation, confirming the adjustment value.
CAN
Corresponds to the RESET key operation, ignoring the adjustments.
END
Corresponds to the SHIFT key operation, keeping the adjustments.
Note
•
In case of D/A adjustment, change the channel using the CHm command, then carry out DO0
or DO1 command, and the request output using the CDm,n command. After the CDm,n
command is being executed by +5 V or –5 V, make sure to confirm by the ENT command.
•
After the display has been stabilized in the range adjustment mode or external input range
adjustment mode, execute the ENT command.
15.1 Adjustments