10-8
IM 253421-01E
Default Output Format in case Integration Measurement
W1 data
Terminator
Wh1data
Terminator
Ah1data
Terminator
Frequency
,
Elapsed
integration time
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Data
number
Terminator
(The data number will only be output in case of recall)
Note
•
When the frequency is set by either of the following methods, only one value is measured,
and that value will be output.
-
by panel keys: FUNCTION key on display C.
-
by communication command: by the “DC” or “EC” command.
After setting the measurement object of frequency, even changing the display C to something
different than VHz or AHz will not result in changing the object of measurement of frequency.
When selecting the output items yourself and you set a frequency item which is not object of
measurement, “999999.E+03” will be output.
•
The displayed values of V (voltage), A (current), W (active power), VA (apparent power), var
(reactive power), Vpk (voltage peak), and Apk (current peak) while the MAX hold function
(see section 4.8) is enabled will be the maximum values (MAX) that are held. The values
output via communications are also set to the maximum values (MAX) that are held.
10.4 Output Format