14-19
IM 253421-01E
Communication Commands 2 (System of Commands Complying to the IEEE 488.2-1987 Standard)
14
[CONFigure]:CURRent:RANGe
Function
Sets the current range (external sensor input
range), queries the current setting.
Syntax
[CONFigure]:CURRent:RANGe
{<current>|(EXTernal,<voltage>)}
[CONFigure]:CURRent:RANGe?
<current>=5mA to 20A (5m, 10m, 20m, 50m,
100m, 200m, 0.5, 1, 2, 5, 10, 20A)
<voltage>=50mV to 200mV (50, 100, 200mV,
for /EX2 option)
=2.5V to 10V (2.5, 5, 10V, for
/EX1 option)
Example
Setting of current range/query
[CONFIGURE]:CURRENT:RANGE 20A
[CONFIGURE]:CURRENT:RANGE?
→
:CONFIGURE:
CURRENT:RANGE 20.0E+00
Setting of external sensor input range/
query (for /EX2 option)
[CONFIGURE]:CURRENT:RANGE EXTERNAL,50MV
[CONFIGURE]:CURRENT:RANGE?
→
:CONFIGURE:
CURRENT:RANGE EXTERNAL,50.0E-03
[CONFigure]:FILTer
Function
Sets the filter ON/OFF, queries the current
setting.
Syntax
[CONFigure]:FILTer {<Boolean>}
[CONFigure]:FILTer?
Example
[CONFIGURE]:FILTER OFF
[CONFIGURE]:FILTER?
→
:CONFIGURE:FILTER 0
[CONFigure]:MHOLd[:STATe]
Function
Sets/queries the MAX hold function.
Syntax
[CONFigure]:MHOLd[:STATe] {<Boolean>}
[CONFigure]:MHOLd[:STATe]?
Example
[CONFIGURE]:MHOLD:STATE OFF
[CONFIGURE]:MHOLD:STATE?
→
:CONFIGURE:
MHOLD:STATE 0
[CONFigure]:MODE
Function
Sets the measurement mode of current and
voltage, queries the current setting.
Syntax
[CONFigure]:MODE {RMS|VMEan|DC}
[CONFigure]:MODE?
Example
[CONFIGURE]:MODE RMS
[CONFIGURE]:MODE?
→
:CONFIGURE:
MODE RMS
[CONFigure]:SCALing?
Function
Queries all settings relating to the scaling
function.
Syntax
[CONFigure]:SCALing?
Example
[CONFIGURE]:SCALING?
→
:CONFIGURE:
SCALING:STATE 0;PT:ELEMENT1 1.000E+00;:
CONFIGURE:SCALING:CT:
ELEMENT1 1.000E+00;:
CONFIGURE:SCALING:SFACTOR:
ELEMENT1 1.000E+00
[CONFigure]:SCALing:{PT|CT|SFACtor}?
Function
Queries all scaling constants related to
{voltage|current|power}.
Syntax
[CONFigure]:SCALing:{PT|CT|SFACtor}?
Example
[CONFIGURE]:SCALING:PT?
→
:CONFIGURE:
SCALING:PT:ELEMENT1 1.000E+00
[CONFigure]:SCALing:{PT|CT|SFACtor}:ELEMent<x>
Function
Sets the scaling constant for
{voltage|current|power} of each element,
queries the current setting.
Syntax
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x> {<NRf>}
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>?
<x>= 1
{<NRf>}=0.001 to 9999
Example
[CONFIGURE]:SCALING:PT:ELEMENT1 1.000
[CONFIGURE]:SCALING:PT:ELEMENT1?
→
:
CONFIGURE:SCALING:PT:ELEMENT1 1.000E+00
Description
The values are rounded as follows:
Less than 1.000: Rounded to three decimal places.
1.000 to 9999: Rounded to four significant digits.
[CONFigure]:SCALing[:STATe]
Function
Sets scaling ON/OFF, queries the current
setting.
Syntax
[CONFigure]:SCALing[:STATe] {<Boolean>}
[CONFigure]:SCALing:STATe?
Example
[CONFIGURE]:SCALING:STATE OFF
[CONFIGURE]:SCALING:STATE?
→
:CONFIGURE:
SCALING:STATE 0
[CONFigure]:SYNChronize
Function
Sets/queries the measurement synchronization
source.
Syntax
[CONFigure]:SYNChronize {VOLTage|
CURRent}
[CONFigure]:SYNChronize?
Example
[CONFIGURE]:SYNCHRONIZE VOLTAGE
[CONFIGURE]:SYNCHRONIZE?
→
:CONFIGURE:
SYNCHRONIZE VOLTAGE
[CONFigure]:VOLTage?
Function
Queries all settings relating to voltage range.
Syntax
[CONFigure]:VOLTage?
Example
[CONFIGURE]:VOLTAGE?Æ:CONFIGURE:VOLTAGE:
RANGE 600.0E+00;AUTO 1
[CONFigure]:VOLTage:AUTO
Function
Sets the voltage auto range ON/OFF, queries
the current setting.
Syntax
[CONFigure]:VOLTage:AUTO {<Boolean>}
[CONFigure]:VOLTage:AUTO?
Example
[CONFigure]:VOLTage:AUTO ON
[CONFIGURE]:VOLTAGE:AUTO?Æ:CONFIGURE:
VOLTAGE:AUTO 1
14.3 Commands