14-18
IM 253421-01E
CONFigure?
Function
Queries all the settings related to the
measurement conditions.
Syntax
CONFigure?
Example
CONFIGURE?Æ:CONFIGURE:WIRING P1W2;
MODE RMS;VOLTAGE:RANGE 600.0E+00;
AUTO 1;:CONFIGURE:CURRENT:
RANGE 20.0E+00;AUTO 1;ESCALING:
ELEMENT1 50.00E+00;:CONFIGURE:FILTER 0;
SCALING:STATE 0;PT:ELEMENT1 1.000E+00;:
CONFIGURE:SCALING:CT:
ELEMENT1 1.000E+00;:CONFIGURE:SCALING:
SFACTOR:ELEMENT1 1.000E+00;:CONFIGURE:
AVERAGING:STATE 0;TYPE LINEAR,8;
:CONFIGURE:SYNCHRONIZE
CURRENT;MHOLD:STATE 0
[CONFigure]:AVERaging?
Function
Queries all the setting values related to the
averaging function.
Syntax
[CONFigure]:AVERaging?
Example
[CONFIGURE]:AVERAGING?
→
:CONFIGURE:
AVERAGING:STATE 0;TYPE LINEAR,8
[CONFigure]:AVERaging[:STATe]
Function
Sets averaging ON/OFF, or queries the current
status.
Syntax
[CONFigure]:AVERaging[:STATe]
{<Boolean>}
[CONFigure]:AVERaging:STATe?
Example
[CONFIGURE]:AVERAGING:STATE OFF
[CONFIGURE]:AVERAGING:STATE?
→
:CONFIGURE:AVERAGING:STATE 0
[CONFigure]:AVERaging:TYPE
Function
Sets the averaging type and constant, queries
the current setting.
Syntax
[CONFigure]:AVERaging:TYPE {LINear|
EXPonent},{<NRf>}
[CONFigure]:AVERaging:TYPE?
{<NRf>}=8, 16, 32, 64 (averaging
constant)
Example
[CONFIGURE]:AVERAGING:TYPE LINEAR,8
[CONFIGURE]:AVERAGING:TYPE?
→
:CONFIGURE:
AVERAGING:TYPE LINEAR,8
[CONFigure]:CURRent?
Function
Queries all setting values relating to the current
range (external sensor range)
Syntax
[CONFigure]:CURRent?
Example
[CONFIGURE]:CURRENT?
→
:CONFIGURE:
CURRENT:RANGE 20.0E+00;AUTO 1;ESCALING:
ELEMENT1 50.00E+00
[CONFigure]:CURRent:AUTO
Function
Sets the current auto range ON/OFF, or queries
the current setting.
Syntax
[CONFigure]:CURRent:AUTO {<Boolean>}
[CONFigure]:CURRent:AUTO?
Example
[CONFIGURE]:CURRENT:AUTO ON
[CONFIGURE]:CURRENT:AUTO?
→
:CONFIGURE:
CURRENT:AUTO 1
[CONFigure]:CURRent:ESCaling?
Function
Queries all scaling constants for the external
sensor.
Syntax
[CONFigure]:CURRent:ESCaling?
Example
[CONFIGURE]:CURRENT:ESCALING?
→
:CONFIGURE:CURRENT:ESCALING:
ELEMENT 50.00E+00
[CONFigure]:CURRent:ESCaling:ELEMent<x>
Function
Sets the scaling constants for the external
sensor for each element separately, queries the
current setting.
Syntax
[CONFigure]:CURRent:ESCaling:
ELEMent<x> {<NRf>}
[CONFigure]:CURRent:ESCaling:ELEMent<x>?
<x>= 1
{<NRf>}=0.001 to 1000
Example
[CONFIGURE]:CURRENT:ESCALING:
ELEMENT1 50.00
[CONFIGURE]:CURRENT:ESCALING:ELEMENT1?
→
:CONFIGURE:CURRENT:ESCALING:
ELEMENT1 50.00E+00
Description
The values are rounded as follows:
Less than 1.000: Rounded to three decimal places.
1.000 to 1000: Rounded to four significant digits.
AVERaging
:
STAT e
<Space>
OFF
ON
<NRf>
?
TYPE
<Space>
LINear
,
<NRf>
EXPonent
?
?
?
;
SYNChronize
<Space>
VOLTage
CURRent
?
14.3 Commands