9-6
IM WT1801-01EN
Resetting Integration (Reset)
The integration time and integrated value are reset. The STOP indicator turns off. The displays for integration-
related functions change to “-----------” (no data).
When an integration error occurs, the START and STOP indicators blink, and the integration status indication
is “Error.”
Holding Integration and Releasing the Hold
Holding Integration
When you press the HOLD key, it illuminates, and the display and communication output of the integration result
are held. Integration continues regardless of whether the display is held.
Releasing the Hold on Integration
If you press the HOLD key when integration is held, the HOLD key light turns off, and the numeric data display
is updated. While the display is held, you can update the display by pressing SINGLE to execute a single
measurement.
The relationship between the hold function and the start and stop operations is as follows:
• If you start integration while the display is held, the display and communication output do not change. When
you release the hold function (turn it off) or make a single measurement by pressing SINGLE, the integration
result at that point is displayed and transmitted.
ON
HOLD
OFF
Displayed value
Integration time
(The dotted line is the integrated value.)
Start
Stop
Reset
• If you stop integration while the display is held, the displayed and transmitted values do not change. When
you release the hold function (turn it off) or make a single measurement by pressing SINGLE, the integration
result at the point when integration was stopped is displayed and transmitted.
Integration time
(The dotted line is the integrated value.)
SINGLE
(Single measurement)
OFF
ON
HOLD
Execution Execution
Execution
Displayed value
Start
Stop
Reset
9 Integrated Power (Watt hours)