When the Storage Mode Is Integ Sync and the Integration Mode Is Continuous or R-Continuous
Storage is performed at the first data update after the storage interval passes. Storage continues even when
integration is reset by the integration timer. When the integration timer is reset, the storage interval timer is also
reset. The figure below is an example for when the integration timer is set to 20 s.
Time (s)
Data
update
Storage
Time (s) 0 5 10 15 20 5 10 15
0 7 14 20 7 14
• When the storage of numeric data at store start (Store At Start) is enabled, storage starts when the time is
at 0 s (store start).
• Storage is performed when the integration mode is continuous integration mode or real-time continuous
integration mode, the integration timer time elapses, and the integrated values are reset. In the above
example, the integration timer is set to 20 s, so the data is stored after 20 s have passed, and then the
integrated values are reset.
Operations when Values Are Held
If you press HOLD after storage has been started, the displayed values are held, and storage is performed as
indicated below.
• When the storage mode is set to Manual and the storage interval is 00:00:00, when the storage mode is set
to Real Time and the storage interval is 00:00:00, or when Storage Mode is set to Event
The storage operation stops, but integration does not.
• When the storage mode is set to Single Shot
The held display values are stored. Even during integration, the held display values are stored.
• In All Other Cases
The held display values are stored. During integration, the values being measured are stored.
• Media access indication icon only illuminates when the WT1800 creates files at the start of storage and
closes them at the end of storage, but the WT1800 is actually always accessing the storage medium.
During the period between when the files are created and when they are closed, do not remove the USB
memory or turn off the power. Doing so may damage the storage medium and corrupt its data.
• When a point has no numeric data, NAN, OL, OF, ERROR, or a blank space will be stored instead. The
places that contain blank spaces are in measurement functions such as the ones for the 0th (DC) and 1st
harmonic orders of ΦU and ΦI, where the screen display is also blank.
• Functions ΔU1 to ΔPΣ are stored according to the delta computation (option) type that you selected in the
delta computation settings.
• When the WT1800 automatically changes the range during storage, because measured data cannot
be acquired, the store operation is paused and no measured data is stored. After the WT1800 finishes
changing the range, storage resumes.
• You cannot use the auto print feature during storage.