17-2
IM WT1801-01EN
Storage Control (Control Settings)
Storage Mode (Store Mode)
You can choose from one of the following methods for starting and stopping storage.
• Manual (Manual storage mode)
When you press STORE START, numeric data is stored at the storage interval for the number of times
specified by the storage count.
• Real Time (Real-time storage mode)
Storage starts after you press STORE START and the scheduled storage start time is reached. You can store
numeric data at the storage interval until the storage stop time (or just the storage count) is reached.
• Integ Sync (Integration-synchronized storage mode)
• Storage starts after you press STORE START and integration starts. You can store numeric data at the
storage interval until integration stops (or just until the storage count is reached) is reached.
• Storage continues even when integration is reset by the
. When the integration timer is
reset, the storage interval timer is also reset.
When independent integration is enabled, storage cannot be performed in integration-synchronized storage
mode.
• Event (Event-synchronized storage mode)
Storage starts after you press STORE START, the measured data is updated, and a user-defined event
occurs. You can store numeric data each time the measured data is updated until the storage count is
reached.
• Single Shot (Single-shot storage mode)
Numeric data is stored whenever you press STORE START. You can store numeric data until the storage
count is reached.
Storage Count (Store Count)
• The storage count can be set to infinity or to a value from 1 to 9999999.
• When the storage count is set to infinity, it is set to 9999999.
• Storage will stop before the specified storage count if the storage destination runs out of available memory or
the stored data size exceeds the maximum value (1 GB).
Viewing and Optimizing the Maximum Storage Count (Optimize Count)
• The WT1800 computes the maximum storage count for the storage destination according to the number of
data types specified in the
.
• You can set the storage count to a value from 0 to the computed maximum storage count.
• You can select Set to set the maximum storage count to the current storage count. However, you cannot do
this if the current storage count is 0.
17 Data Storage