17-7
IM WT1801-01EN
Stopping Storage (STORE STOP)
• You can temporarily stop storage by pressing STORE STOP.
• When storage stops, the STORE STOP key blinks, and “Store:Stop” appears in the upper left of the screen.
• When the storage count is zero and the WT1800 is in a storage-ready state (Ready)
*
and you press STORE
STOP, storage is reset. The storage data file (.wts) and storage header file (.hds) are deleted.
* This state can occur in real-time storage mode when you have pressed STORE START but the scheduled
storage start time has not yet arrived (the storage operation has not begun).
Ending Storage
• Depending on the storage mode, storage automatically stops and the storage state changes to “Cmpl” under
one of the following conditions.
• Manual storage mode (Manual)
Storage continues until the storage count is reached.
• Real-time storage mode (Real Time)
Storage continues until the storage count or the scheduled storage stop time.
• Integration-synchronized storage mode (Integ Sync)
Storage continues until the storage count is reached, and then storage is completed (Cmpl). When
integration stops, the following occurs.
• If integration cannot be restarted without being reset, the storage state changes to “Stop.”
• If integration can be restarted without being reset, the storage state changes to “Ready.”
• Event-synchronized storage mode (Event)
Storage continues until the storage count is reached.
• Single-shot storage mode (Single Shot)
Storage continues until the storage count is reached.
• After storage is completed, the following occurs.
• The STORE STOP key illuminates, and “Store:Cmpl” appears in the upper left of the screen.
• Writing to the stored data file (.wts) and header file (.hds) finishes, and the files are closed.
• When automatic CSV conversion (Auto CSV Conversion) is enabled, an ASCII format stored data file (.csv)
is created.
Storage will stop before the specified storage count if the storage destination runs out of available memory or
the stored data size exceeds the maximum value (1 GB).
Resetting Storage (STORE RESET)
• When you reset storage, the WT1800 enters into the reset store state, in which the store state indicator is
blank.
• If you reset storage when the store state is “Stop,” writing to the stored data file (.wts) and header file (.hds)
finishes, and the files are closed. When automatic CSV conversion (Auto CSV Conversion) is enabled, an
ASCII format stored data file (.csv) is created.
• When the store state is “Cmpl,” the stored data file (.wts) and header file (.hds) are already closed, so no file
operations are performed when storage is reset.
17 Data Storage