4-7
IM WT1801-01EN
The pulse input range setting applies to input signals whose type has been set to pulse. For input signals
whose type has been set to analog, there is no need to configure this setting.
Revolution signal pulses per revolution (Pulse N)
Set the number of pulses per revolution to a value from 1 to 9999.
The following equation is used to compute the rotating speed.
X
N
Rotating speed = S
−
NULL
S:
The scaling factor
*
X:
The number of input pulses from the revolution sensor per minute
N:
The number of pulses per revolution
NULL:
* When the scaling factor is 1, the rotating speed is the number of revolutions per minute (min
–1
or rpm). If
the revolution signal is a changed signal, you can set the
to determine the rotating speed
before the change.
The setting specifying the number of revolution signal pulses per revolution applies to input signals whose
type has been set to pulse. For input signals whose type has been set to analog, there is no need to configure
this setting.
Synchronous Speed (Sync Speed)
Motor Poles (Pole)
Set the number of motor poles to a value from 1 to 99. This number is used to compute the sync speed.
Frequency Measurement Source (Source)
• Select the frequency measurement source to use to compute the sync speed from the options below. The
frequency measurement source is used to compute the sync speed. The available options vary depending on
the installed elements.
U1, I1, U2, I2, U3, I3, U4, I4, U5, I5, U6, and I6
• On models that do not have the add-on frequency measurement option installed, select a voltage or current
whose frequency is being measured (
voltage/current to measure the frequency of
current whose frequency is not being measured, an error will occur.
• Normally, you should select the voltage or current that is being applied to the motor and whose frequency is
being measured. If you select the frequency of a voltage or current that is not being applied to the motor, the
sync speed may not be determined properly.
Sync Speed Equation
The sync speed unit is fixed at min
–1
or rpm.
120 × Fs
Pole
Sync speed (min
−
1
) =
Fs:
The frequency of the frequency measurement source (Hz)
Pole:
The number of motor poles
4 Motor Evaluation Conditions (Option)