9-1
IM WT1801-01EN
9 Integrated Power (Watt hours)
The WT1800 can integrate the active power (watt hour), the current (ampere hour), the apparent power (volt-
ampere hour), and the reactive power (var hour) values.
For a list of the measurement functions related to integrated power (watt hours) and their descriptions, see “Items
That This Instrument Can Measure.”
You can set the following items.
•
Enabling or Disabling Independent Integration (Independent Control)
•
Independently integrated elements (Element Object)
•
Starting, Stopping, and Resetting Integration (Start/Stop/Reset)
•
Integration Conditions (Integ Set)
Integration-Related Display Indications
Integration status
Reset
Displayed when the integrated value is reset and integration can be started.
Start
Displayed along with the elapsed integration time when integration has been started.
Stop
Displayed along with the elapsed integration time when integration has been stopped.
• When integration has been stopped forcibly through the pressing of the Stop soft key, “Stop” appears in
yellow. You can press Start to resume integration.
• In real-time normal integration mode and real-time continuous integration mode, once the scheduled
integration stop time is reached, “Stop” appears in orange. Even if you press Start, integration is not resumed.
To start integration, you must first reset it.
Ready
In real-time normal integration mode, when you press Start, before the scheduled integration start time is
reached, “Ready” and the scheduled integration start time appear.
Scheduled integration start time
TimeUp
After the specified integration timer time elapses, integration stops automatically, and “TimeUp” and the elapsed
integration time appear.
Error
In the following situations, integration stops automatically, and “Error” and the elapsed integration time appear.
• The integration time reaches the maximum integration time (10000 hours).
• The integrated value reaches its maximum or minimum displayable value.
• A power failure occurs when integration is in progress. Even when a power failure occurs, the WT1800 stores
and holds the integration result. When the power returns, integration is stopped, and the integration result
calculated up to the point when the power failure occurred is displayed.
Example of the Display When the Integration Time Is Set to a Non-Zero Value
Elapsed integration time
Integration timer