17-4
IM WT1801-01EN
Stored Items (Item Settings)
You can set which numeric data items to store. You can choose to store the displayed numeric items (Displayed
Numeric Items) or the selected items (Selected Items).
Displayed Numeric Items (Displayed Numeric Items)
The numeric data items displayed on the screen are stored. The stored items vary as indicated below depending
on the display.
• When Numeric Values Are Displayed in the 4-, 8-, or 16-Value Display or the Matrix Display
All the measurement functions on the page that is displayed when storage starts are stored in the order that
they are displayed.
• When Numeric Values Are Displayed in the Single or Dual Harmonics List
In addition to the data described above, the data of harmonics that are not displayed on the screen is stored
up to the maximum measurable order (Max Order).
• When Numeric Values Are Displayed in the All Items or Custom Display
Storage cannot be performed. An error message is displayed at store start.
• In Non-Numeric Displays (Waveform display, trend display, etc.)
The WT1800 stores data according to the numeric display settings. For example, when the waveform
display is shown, if the 16-value display would appear when you pressed NUMERIC, then the data for the
measurement functions on the page of the 16-value display that would appear is stored.
During storage, even if you change the numeric data display format or the displayed items, the data is stored
according to the numeric data that was displayed at store start.
Selected Items (Selected Items)
You can select the types of numeric data to store. To select which types of numeric data to store, click Items.
Items (Items)
If you set the numeric data items to store to Selected Items, you can select which types of numeric data to store.
Element (Element)
To select an element or wiring unit that you want to store the data of, select or clear its check box. You can select
from the following options.
Element1, Element2, Element3, Element4, Element5, Element6, ΣA, ΣB, and ΣC
Function (Function)
Select whether to store the data of a function by selecting or clearing its check box. You can select any of the
measurement function types listed under “Items That This Instrument Can Measure.”
Selecting All Functions (All ON)
The data of all measurement functions is stored.
Deselecting All Functions (All OFF)
None of the data for any of the measurement functions is stored.
Preset 1 (Preset 1)
The data of the following measurement functions is stored for all elements and wiring units.
*
Urms, Irms, FreqU, FreqI, P, S, Q, λ, and Φ
Preset 2 (Preset 2)
The data of the following measurement functions is stored for all elements and wiring units.
*
WP, WP+, WP–, q, q+, q–, Time, WS, and WQ
* If the wiring system setting (Wiring) is configured so that a wiring unit does not exist, the data for the functions
of tha
t wiring unit is not stored. For example, if ΣC does not exist, the data for ΣC is not stored.
17 Data Storage