![YOKOGAWA VZ20X User Manual Download Page 89](http://html1.mh-extra.com/html/yokogawa/vz20x/vz20x_user-manual_3498955089.webp)
7-22
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
●
Measured value, measured value time stamp, device status, measured value
status
The registers for storing the measured value, measured value time stamp, device status,
and measured value status are shown below.
The values of these registers are updated according to the data acquisition interval.
Data acquisition interval
Register update interval
1 ms
10 ms
10 ms
10 ms
50 ms
50 ms
100 ms
100 ms
Two types of register assignments are available to simplify accessing of accessing
registers.
・
Continuous reading of measured values (Table 1): Placement of registers by which
measured values and statuses are respectively continuous.
・
Reading of measured values and statuses by individual channel (Table 2):
Placement of registers by which measured values and statuses can be acquired as
a set.
Table 1. Continuous reading of measured values
Register number
(Reference no.)
Hex number
(hexadecimal)
Register name
R/W
Descriptions
40101
0064
Measured value time stamp
R
The time of measured values
can be read.
This setting is common to all
channels.
The coordinated universal time
or UTC can be read as a 64-bit
unsigned integer.
40102
0065
R
40103
0066
R
40104
0067
R
40105
0068
Device status
R
The device status of the VZ20X
can be read.
40106
0069
CH1 measured value
R
Measured values can be read
as 32-bit signed integers.
The range changes depending
on the input range.
40107
006A
R
40108
006B
CH2 measured value
R
40109
006C
R
40110
006D
CH3 measured value
R
40111
006E
R
40112
006F
CH4 measured value
R
40113
0070
R
40114
0071
CH5 measured value
R
40115
0072
R
40116
0073
CH6 measured value
R
40117
0074
R
40118
0075
CH7 measured value
R
40119
0076
R
40120
0077
CH8 measured value
R
40121
0078
R
40122
0079
CH1 measured value status
R
The status of measured values
can be read.
40123
007A
CH2 measured value status
R
40124
007B
CH3 measured value status
R
40125
007C
CH4 measured value status
R
40126
007D
CH5 measured value status
R
40127
007E
CH6 measured value status
R
40128
007F
CH7 measured value status
R
40129
0080
CH8 measured value status
R
Summary of Contents for VZ20X
Page 16: ...Contents 3 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Revision Information 1...
Page 17: ...Contents 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 23: ...2 2 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 29: ...3 6 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 145: ...9 10 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 149: ...App 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...