Acquisition of time
・
GA10
The time is acquired automatically only once from the GA10 when the connection is
made.
Set the setting value of GA10 setting item "Data time" to "Device time".
・
Other than GA10 (PC/PLC)
Write the time of the PC/PLC via Modbus/TCP communication after the VZ20X is
started up.
Write the time also when the power supply is turned off midway.
At writing of the time, the time stamp is omitted or set in duplicate.
Device targeted for writing of time: VZ20X, Connection type setting value: 1 unit,
master unit of daisy chain
PC/PLC
VZ20X
VZ20X power supply on
The date and time are acquired at
startup.From here on, the
continuity of the time stamp on
the VZ20X is held as the date and
time are not received from the
PC/PLC.