![YOKOGAWA VZ20X User Manual Download Page 114](http://html1.mh-extra.com/html/yokogawa/vz20x/vz20x_user-manual_3498955114.webp)
7-47
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
Continued
PDU
1
1
1
1
1
1
1
1
10
CH1
STS1
CH2
STS1
CH3
STS1
CH4
STS1
CH5
STS1
CH6
STS1
CH7
STS1
CH8
STS1
Year/month
/day 2
....
PDU
4
4
1
1
1
1
1
1
1
1
CH7
Data4
CH8
Data4
CH1
STS4
CH2
STS4
CH3
STS4
CH4
STS4
CH5
STS4
CH6
STS4
CH7
STS4
CH8
STS4
Continued
PDU
2
UNIT status
Data is arranged in the following order when it is acquired: data acquisition status,
year/month/day 1, time 1, CH1-Data1, CH2-Data1, CH3-Data1, CH4-Data1, CH5-Data1,
CH6-Data1, CH7-Data1, CH8-Data1, CH1-STS1, CH2-STS1, CH3-STS1, CH4-STS1,
CH5-STS1, CH6-STS1, CH7-STS1, CH8-STS1, year/month/day 2, time 2
,
・・・・・・
, UNIT
status
.
*Data: Measured value
Measured values are time-stamped. The format of the time stamp is as follows:
Year/month/day: yyyy-mm-dd
Time: hh:mm:ss.nnn
●
Response (during an error)
Element
MBAP Header
PDU
Number of
bytes
2
2
2
1
1
1
Command
elements
Transfer
ID
Protocol
ID
Number
of bytes
Unit ID Function
code
(0x47)
Data
acquisition
status
(Hex)
Arbitrary
0000
0003
01
47
01 or 02
■
Details of error code
Error Code
Meaning
Cause
01
Function code error
The function code does not exist.
02
Register
A No. out of range was specified.
03
Register
A number of registers out of range was specified.
Summary of Contents for VZ20X
Page 16: ...Contents 3 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Revision Information 1...
Page 17: ...Contents 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 23: ...2 2 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 29: ...3 6 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 145: ...9 10 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 149: ...App 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...