8-6
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
The following table summarizes the ranges that can be adjusted.
Range
Types of Input That Use Adjustment Values
20 mV
DC voltage 20mV, thermocouples T, B, S, R
60 mV
DC voltage 60mV, thermocouples K, J, N, E, C
200 mV
DC voltage 200mV
1 V
DC voltage 1V
2 V
DC voltage 2V
、
Standard signal 0.4-2V
6 V
DC voltage 6V
、
Standard signal 1-5V
20 V
DC voltage 20V
60 V
DC voltage 60V
200Ω
-3 wire
Resistance temperature detector Pt100-H 3-wire
400Ω
-3 wire
Resistance temperature detector Pt100 3-wire
200Ω
-4 wire
4-wire resistance 200
Ω,
Resistance temperature detector Pt100-H 4-wire
400Ω
-4 wire
Resistance temperature detector Pt100 4-wire
2000Ω
-4 wire
4-wire resistance 2000
Ω
■
Required equipment
For details on the equipment required for adjustment, check "Required equipment" at
"8.1 Input Calibration."