![YOKOGAWA VZ20X User Manual Download Page 112](http://html1.mh-extra.com/html/yokogawa/vz20x/vz20x_user-manual_3498955112.webp)
7-45
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
■
71 Acquire data
This command can acquire "4 data acquisition intervals' worth of data x data of 8
channels".
●
Function
・
Acquisition of latest data acquisition value
・
Acquisition of data previous access value (the same data is acquired again when a
communication error occurs at acquisition of the latest value)
●
Function selection
Function selection
code
Description
0001
Acquisition of latest data acquisition value
0002
Acquisition of data previous access value
The response is the same whichever function selection code is set.
●
Data acquisition error status
Error Code
Description
00
No error
01
Buffer is not full.
02
Data omission occurred.
●
CHn STS (n=1 to 8)
This is the measured value status of each input channel. This displays the error
information of measured values as a single byte. The 0 to 5th bits are handled as
numerical values. The 6th and 7th bits express errors in bits. Error status is expressed in
the numerical values of the 0 to 5th bits as the logical sum of the 6th and 7th bits.
When +overscale, measured value error and RJC error occur simultaneously, the values
are as follows:
+overscale: 1
Adjustment value error: 64
RJC error: 128
Error status = 1 (+overscale) + 64 (Adjustment value error) + 128 (RJC error) = 193
Summary of Contents for VZ20X
Page 16: ...Contents 3 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Revision Information 1...
Page 17: ...Contents 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 23: ...2 2 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 29: ...3 6 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 145: ...9 10 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 149: ...App 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...