![YOKOGAWA VZ20X User Manual Download Page 146](http://html1.mh-extra.com/html/yokogawa/vz20x/vz20x_user-manual_3498955146.webp)
App-1
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
Appendix.
Appendix.1 List of Parameters
Setting > Communication
Parameter Name
Setting Range
Default
Explanation
Tag
Max. 16, 2- or 1-bytes
characters
(alphanumeric characters,
symbols and Chinese
character)
VZ
Sets the tag name of the VZ20X.
User ID
(*1)
Max. 16, 1-byte characters
(alphanumeric characters
and symbols)
USER1
Sets the user ID.
Password
(*1)
Max. 16, 1-byte characters
(alphanumeric characters
and symbols)
USER1
Sets the password.
Ethernet Port IP
address
(*1)
0.0.0.0 to 255.255.255.255 192.168.1.1
Sets the IP address of the VZ20X.
Subnet mask
(*1)
0.0.0.0 to 255.255.255.255 255.255.255.0 Sets the subnet mask of the VZ20X.
Default
gateway
(*1)
0.0.0.0 to 255.255.255.255 0.0.0.0
Sets the default gateway of the VZ20X.
Connection
permission
(*1)
OFF, IP address, MAC
address
OFF
Sets the connection permission of the
device.
Connection permission can be set on up
to two devices.
Connection
permission IP
address 1
(*1)
0.0.0.0 to 255.255.255.255 0.0.0.0
Sets the connection permission IP
address 1.
Connection
permission IP
address2
(*1)
0.0.0.0 to 255.255.255.255 0.0.0.0
Sets the connection permission IP
address 2.
Connection
permission MAC
address1
(*1)
00-00-00-00-00-00 to FF-FF-
FF-FF-FF-FF
00-00-00-00-
00-00
Sets the connection permission MAC
address 1.
Connection
permission MAC
address2
(*1)
00-00-00-00-00-00 to FF-FF-
FF-FF-FF-FF
00-00-00-00-
00-00
Sets the connection permission MAC
address 2.
DHCP
(*1)
OFF, ON
ON
Sets the DHCP function.
Use this function only at initial setup.
During actual use, set DHCP to OFF and
fix the IP address.
Ethernet
connection port
number
Fixed to 502
502
The port number is used for
communication with the VZ configurator
and for communication using
Modbus/TCP function codes 03 and 16.
Firmware update
permission via
Ethernet
OFF, ON
OFF
Sets permission to update firmware over
Ethernet.
Setting > USB Type-C Detection
Parameter Name
Setting Range
Default
Explanation
USB Type-C
detection
Enable, Disable
Enable
Sets USB Type-C detection. This
function is enabled with firmware version
R1.02.01 or later.
*1: Communication is turned off, and these parameters require beginning with a new connection when setting values
were changed in a USB connection.
Summary of Contents for VZ20X
Page 16: ...Contents 3 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Revision Information 1...
Page 17: ...Contents 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 23: ...2 2 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 29: ...3 6 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 145: ...9 10 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 149: ...App 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...