![YOKOGAWA VZ20X User Manual Download Page 69](http://html1.mh-extra.com/html/yokogawa/vz20x/vz20x_user-manual_3498955069.webp)
7-2
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
7.1.1
Input Type
The following table summarizes the types and ranges of the measurement inputs (DC
voltage, standard signal, 4-wire resistance, thermocouple, 3-wire resistance temperature
detector and 4-wire resistance temperature detector) to the VZ20X.
Input type
Measurement range
Unit
Lower limit value
Upper limit value
OFF (default)
60 V
-60.00
60.00
V
DC voltage
20 mV
-20.000
20.000
mV
60 mV
-60.00
60.00
mV
200 mV
-200.00
200.00
mV
1 V
-1.0000
1.0000
V
2 V
-2.0000
2.0000
V
6 V
-6.000
6.000
V
20 V
-20.000
20.000
V
60 V
-60.00
60.00
V
Standard signal
0.4 to 2 V
0.4000
2.0000
V
1 to 5 V
1.0000
5.0000
V
4-wire resistance
200
Ω
4-wire
0.00
200.00
Ω
2000
Ω
4-wire
0.0
2000.0
Ω
Thermocouple
R
0.0
1760.0
°C
S
0.0
1760.0
°C
B
0.0
1820.0
°C
K
-270.0
1370.0
°C
E
-270.0
800.0
°C
J
-200.0
1000.0
°C
T
-270.0
400.0
°C
N
-270.0
1300.0
°C
C
0.0
2315.0
°C
3-wire resistance
temperature
detector
Pt100 3-wire
-200.0
850.0
°C
Pt100-H (3-wire) -100.00
100.00
°C
4-wire resistance
temperature
detector
Pt100 4-wire
-200.0
850.0
°C
Pt100-H (4-wire) -100.00
100.00
°C
7.1.2
Temperature Unit
When the input type is a thermocouple or resistance temperature detector, the
temperature unit used is °C.
Parameter name
Contents
Temperature Unit
°C
Summary of Contents for VZ20X
Page 16: ...Contents 3 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Revision Information 1...
Page 17: ...Contents 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 23: ...2 2 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 29: ...3 6 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 145: ...9 10 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 149: ...App 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...