![YOKOGAWA VZ20X User Manual Download Page 122](http://html1.mh-extra.com/html/yokogawa/vz20x/vz20x_user-manual_3498955122.webp)
8-7
IM 77V01B01-01EN
5th Edition: 2023.03.07-00
■
Adjustment procedure
1
Wire the VZ20X to the calibration equipment as shown in the figure below, and
sufficiently warm up each of the equipment. (Warm up the VZ20X for at least 40
minutes.)
For precautions in wiring, see "Chapter 4 Wiring."
・
Wiring for DC voltage range
・
Wiring for 3-wire resistance temperature detector range
・
Wiring 4-wire resistance temperature detector/4-wire resistance range
2
Check that the ambient temperature, humidity and other operating environment
conditions are within specification (23±2°C, 55±10% RH).
Zero-point adjustment
Full-point adjustment
1
2
3
4
VZ20X
input terminal
1
2
3
4
DC voltage
generator
DC V
Output
VZ20X
input terminal
HI
LO
Short-circuit input terminal
1
2
3
4
VZ20X
input terminal
A
B
b
1
2
3
4
A
B
b
Reistor
Output
VZ20X
input terminal
* Ensure that the resistance of 3 conductors is the same value.
HI
LO
Zero-point adjustment
Full-point adjustment
Short-circuit input terminal
1
2
3
4
VZ20X
input terminal
a
A
B
b
1
2
3
4
a
A
B
b
Resistor
Output
VZ20X
input terminal
* Ensure that the resistance of 4 conductors is the same value.
HI
LO
Zero-point adjustment
Full-point adjustment
Short-circuit input terminal
Summary of Contents for VZ20X
Page 16: ...Contents 3 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Revision Information 1...
Page 17: ...Contents 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 23: ...2 2 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 29: ...3 6 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 145: ...9 10 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...
Page 149: ...App 4 IM 77V01B01 01EN 5th Edition 2023 03 07 00 Blank Page...