<9. Troubleshooting>
9-1
IM 12K01B02-01E
9
Troubleshooting
9.1
Table of Status and Operation
Item
Status
LED
(POWER/ALARM)
: green light on
: green blinking
Contact
(batch alarm)
Contact
(maintenance)
Output status
Alarm
display
Alarm
history
Common
MEAS.
Instantaneous value output
MAINTENANCE
Maintenance output
Maintenance
CALIBRATION
Maintenance output
Maintenance
DATA
Instantaneous value output
9.2
Table of alarm and Operation
Item
Status
LED (POWER/ALARM)
: green light on
: red light on
: green blinking
: red blinking
Contact
(batch
alarm)
Contact
(analyzer
alarm)
Output status
Alarm
display
Alarm
history
Common
WARM-UP
Warm-up output
CELL MOTOR ERR.
Alarm output
SENSOR COMM. ERR.
*1
Alarm output
SAMPLE LACK
Instantaneous
value output
PIO COMM. ERR.
*2
Alarm output
POWER ON/OFF
Instantaneous
value output
BATTERY ERR.
Instantaneous
value output
LAMP ERR.
Alarm output
UV
ZERO CAL.
Instantaneous
value output
SPAN CAL.
Instantaneous
value output
VIS
ZERO CAL.
Instantaneous
value output
SPAN CAL.
Instantaneous
value output
COD
HIGH
Instantaneous
value output
LOW
Instantaneous
value output
TURB
(turbidity)
HIGH
Instantaneous
value output
LOW
Instantaneous
value output
SPAN CAL.
Instantaneous
value output
*1
SENSOR COMM.ERR: Communication error between detector substrate and
converter substrate
*2
PIO COMM.ERR: Communication error between converter CPU substrate and
input/output substrate