<5. Calibration>
5-18
IM 12K01B02-01E
8. Enter the concentration value of compounded TURB calibration solution with numeric
keypad, and press [SET],
Input range : 0.00 to 9999.99
The screen returns to the TURB. CAL. screen.
NOTE
[CLEAR]: Clears all input numeric value.
[BACK]: Clears the rightmost number.
CAUTION
The decimal place of input range varies according to the setting in “6.5.2 setting of Decimal place
for each Component”
CAUTION
Enter the concentration value of compounded calibration solution for TURB calibration value.
9. After waiting for 3 minutes and confirming that the measurement value is stabilized, press
[SPAN].
The “Start span cal.?” screen is displayed.
10. Press [YES] to start the span calibration.
“Calibrating” screen is displayed.
After calibration, “Calibration finished” screen is displayed.
11. Press [YES].
The span calibration value for TURB (turbidity) is updated.
12. The TURB span calibration is finished.
Set the detector in the measuring tank.
CAUTION
Do not reuse the calibration solution used in the calibration tank. The accuracy of calibrated value
may be lost.