<6. Functions>
6-7
IM 12K01B02-01E
NOTE
If the setting is changed while holding the readings, the setting change is enabled after the hold is
released.
6.4.4
Output Condition Setting
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[SETTING]
Refer to “6.4.2 Signal Input Setting”
OUTPUT CONDITION
SETTING screen
SETTING 1/2
screen
[OUTPUT CONDITION SETTING]
Here enables the output conditions setting (such as the alarm discrimination time and the minus
display).
For normal usage, default setting can be used (no additional setting is required).
1. Press [OUTPUT CONDITION SETTING] on the SETTING 1/2 screen.
The OUTPUT CONDITION SETTING screen appears.
[OUTPUT CONDITION
SETTING]
2. Press the button for ALARM DISCRIMI. TIME to change the discrimination time of the
alarm output, and enter the value of setting range. Press the button for MINUS DISPLAY to
activate/deactivate the minus display.
Setting item
Description
Initial setting
ALARM
DISCRIMI. TIME
Sets the alarm delay time to ignore the spike alarm.
When the time exceeds the ALARM DISCRIMI. TIME, the alarm is
triggered and sounds continuously.
(Set to "0" minute if you want to occur the alarm immediately.)
Setting range: 0 to 99 minuites
5 min
MINUS
DISPLAY
Sets whether to display the measurement value.
Setting range: OFF or ON
OFF
The screen returns back to the OUTPUT CONDITION SETTING screen.
The output condition setting is complete.
6.4.5
Time Adjustment
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[SETTING]
Refer to “6.3 Maintenance Screen Display”
SETTING 1/2 screen
SETTING 2/2 screen
[ ]
The clock can be set here.
1. Press the [TIME ADJUSTMENT] button on the SETTING 2/2 screen.
The TIME ADJUSTMENT screen appears.