<8. Maintenance>
8-12
IM 12K01B02-01E
8.3.2
Removal of Inner Tank of Measuring Tank
The removal of the inner tank of the measuring tank is described here.
1. Stop the supply of sample water and turn the power OFF.
2. Remove the detector and blind plate.
3. Hold the inner tank of the measuring tank by both hands and turn counterclockwise to
remove it.
Inner tank of the measuring tank
Body of the measuring tank
4. Remove the current plate from the inner tank of the measuring tank. As the spring (leg) of
the current plate protrudes out of the screw part on the lower part of the inner tank, it can be
removed upward by pressing inside.
Inner tank of the measuring tank
Current plate
Spring
5. After cleaning the inner tank of measuring tank, put the current plate back in the original
position.
6. Mount the inner tank to the measuring tank by aligning the screw, and turn clockwise to fix.
Screw part