<8. Maintenance>
8-1
IM 12K01B02-01E
8 Maintenance
Calibration and periodical maintenance are required in order to maintain the normal operation
and performance of the unit.
Regarding calibration, refer to “5 Calibration”
8.1
Maintenance item
Maintenance item
Recommended Maintenance cycle
Referece of Time
to clean/replace
Subject
Description
Weekly Monthly
Every
half year
Yearly
Every
3 years Accordingly
Analyzer
unit
Cleaning of measuring cell
When dirt can be seen.
Inspection and replacement
of wipers
When dirt can be seen.
Replacement of desiccant
agent in analyzing case
Periodical replacement
(every half year)
Replacement of desiccant
agent in measuring cell
Periodical replacement
(every half year)
Replacement of inner parts
When the measurement
value is abnormal
Replacement of UV lamp
Periodical replacement (yearly)
When the measurement value
is abnormal
When the UV lamp does not
illuminate
Replacement of motor
Periodical replacement (yearly)
Replacement of seal washers
Periodical replacement (yearly)
Replacement of V-rings
Periodical replacement (yearly)
Replacement of gear
Periodical replacement (yearly)
Replacement of cell
When dirt can be seen
When it is damaged
Replacement of case packing
Periodical replacement (yearly)
Control
unit
Replacement of battery for
clock
Periodical replacement
(every 3 years)
Replacement of substrate and
electric parts
When malfunctioning
Sampling
unit
Cleaning of measuring tank
When dirt can be seen
When clogged
Cleaning of overflow tan
When dirt can be seen
When clogged
Cleaning of piping
When dirt can be seen
When clogged
Cleaning of ball valve
When dirt can be seen
When clogged
The maintenance cycle in the table is the typical interval when using under standard conditions,
and does not guarantee the measurement values. They may vary according to the circumstances
or usage conditions set the proper cycle after using for a certain period of time.
CAUTION
Maintenance items for which the maintenance period is over 1 year are not described in this
manual.
These should be performed by YOKOGAWA service personnel.