<8. Maintenance>
8-8
IM 12K01B02-01E
9. Take the cell plate C and remove the desiccant agent.
Desiccant agent
10. Set the cell plate C, cell packing, and cell plate D as shown in the figure, and tighten the
screw for 2 or 3 threads.
Cell packing
Cell plate C
Cell plate D
Flat washer
Screw
Spring washer
11. Peel off the backing paper of the two-sided tape of the new desiccant agent, and adhere on
the center part of the cell plate C. Before adhesion, brush off the powder on the desiccant
agent.
Desiccant agent
(adhere)
12. Insert this block into the cell, and fully tighten the screw to attach the cell packing firmly on
the inside surface of cell.
Flat washer
Spring washer
CAUTION
• Confirm that no moisture or contamination is adhered inside the cell. If there is moisture or
contamination, thoroughly wipe it off with a soft cloth.
• Tighten the screw after checking that no moisture or contamination is adhered on the seal
part of the cell packing.