<6. Functions>
6-6
IM 12K01B02-01E
1. Press the output item button to be set on the OUTPUT SETTING screen.
• MAINT. OUTPUT
• WARM-UP OUTPUT
• ALARM OUTPUT
The screens for the output of each item appear as follows;
The screen shows an example for MAINT. OUTPUT.
[HOLD]
/[DIRECT]
/[PRESET xxmA]
2. Select the setting item, and press [SET]
Setting item
Description
Initial setting
HOLD
Outputs the adjacent readings while holding.
If there are no adjacent readings when the power is
ON, the latest readings before the power OFF are held.
—
DIRECT
Outputs the measured readings directly
.
—
PRESET
Outputs the preset values.
In this case, the log data is the reading equivalent to
the setting output value.
During maintenance and alarm: 21.0 mA
During warm-up: 4.0 mA
The screen returns back to the OUTPUT SETTING screen.
Output settings during operation are complete.
l
When setting to PRESET:
1. Press the [PRESET] button on the lower right of screen.
The PRESET screen for each item appears.
[PRESET]
2. Enter the preset value with the numeric keypad, and press [SET],
Input range: 0.0 mA to 21.0 mA
CAUTION
When it is preset, the log data also becomes the preset value. The value corresponds to the
following calculating formula.
When the analog output level is 4 mA to 20 mA:
Full scale of measurement value 3 (preset value - 4) / 16
When the analog output level is 0 mA to 16 mA:
Full scale of measurement value 3 preset value / 16