TURB (turbidity) Measurement
TURB measurement is performed by using the visible light (VIS) absorbance.
The measurement method is the transmitting absorption method.
Calibration with the standard solution enables a simple measurement of TURB.
Measure in the measurement range of 200 deg. or below. If measuring in a measurement range
that exceeds the specified value, the linearity deteriorates due to the scattering light influence.
l
How to Calculate the COD Conversion Factor
If it is judged that there is obviously a linear function between the measurement concentration
Xi (i = 1, 2, ...n) and corresponding manual analysis value Yi (i = 1, 2, ...n), the calibration curve
formula (regression formula) is expressed as follows:
Y = a
+
bX
The gradient of this regression line X-axis (b) and the Y-axis segment (b) can be calculated by
using the measurement value and the least-square method as in the following formula.
a=Y
-
bX
b=
=
S
(
X
i
- X
)
2
n
S
X
i
Y
i
-
(
S
X
i
)(
S
Y
i
)
(
X
i
- X
)(
Y
i
- Y
)
n
S
X
i
2
-
(
S
X
i
)
2
“a” and “b” are rounded to a value which is larger by 1 digit according to the number of significant
figures of the measurement value.