<9. Troubleshooting>
9-3
IM 12K01B02-01E
9.4
Causes and Measures for Alarm
Item
Alarm
Causes
Measures
Refer to:
Common
WARM-UP Waiting to be stabilized after the
detector power ON.
No problem. (It takes approximately
60 minutes until the detector is
stabilized.)
"4.2 Starting Operation"
CELL
MOTOR
ERR.
Contamination of measuring cell.
Clean the measuring cell.
"8.2.1 Cleaning Method for
Measuring Cell"
Failure in detector cell-motor.
Replace the motor
.
Service call
SENSOR
COMM.
ERR.
Poor connection of detector cables. Connect the detector cables
properly.
Poor connection of detector cables
or converter, or faulty parts.
Connect the inside cables of the
detector and converter again.
Service call
SAMPLE
LACK
Sample water supply is stopped.
Check the sample water supply and
turn supply on.
"4.1 Preparation for Operation"
Valve shut in improper condition.
V-1 to V-4
"4.1 Preparation for Operation"
Clogging of piping to the overflow
tank.
Clean the piping.
"8.3 Cleaning of Sampling
Unit"
Failure in float switch for no sample
detection.
Replace the float switch
Service call
PIO COMM.
ERR.
Communication error between input/
output substrate and CPU substrate
of converter.
Turn the power OFF/ON. This may
solve the problem.
Poor connection of converter cables,
or faulty parts.
Connect the inside cables of control
unit again.
Service call
POWER
ON/OFF
Power OFF by the electric outage.
No problem.
BATTERY
ERR.
Backup battery for clock has expired. Replace the clock backup battery.
Service call
LAMP ERR.
Contamination of measuring cell.
Clean the measuring cell.
"8.2.1 Cleaning Method for
Measuring Cell"
Light source is temporarily off
Lamp may not be illuminated until
the light source is warmed up. Turn
the power ON/OFF and wait for
approximately 30 minutes as it may
be restored.
Service call
Output is decreased by deterioration
of light source.
Replace the light source.
Service call
Poor connection of converter cables,
or faulty detector or substrate
Connect the inside cables of control
unit again.