<6. Functions>
6-11
IM 12K01B02-01E
[UV]
[COD]
6.5.1
Measuring Range Setting for each Component
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[MEAS.SETTING]
Refer to “6.5 Setting Screen Display for each Measuring Component”
MEAS. SETTING for each
measuring component screen
MEAS. SETTING
screen
[UV][VIS][COD][TURB]
The measuring range for each measuring component is set here.
1. Press the button for MEAS. RANGE on the MEAS. SETTING for each measuring
component screen.
The MEAS. RANGE SETTING for each measuring component screen appears.
The screen shows an example for UV.
MEAS. RANGE
2. Enter the numeric setting range and press [SET].
The screen returns back to the MEAS. SETTING for each measuring component screen.
The measuring range setting for each measuring component is complete.
Perform the same procedure to set the measuring range for the other components.
CAUTION
• Set the measuring range so as not to exceed the full scale of the measurement value (level
of approximately 80%).
• If the measuring range is changed while measuring, the measurement value may go wrong.
In this case, perform the calibration using a matching calibration solution.
• The measuring range for UV-αVIS is the same as that for UV.