<5. Calibration>
5-17
IM 12K01B02-01E
3. After cleaning, pour the TURB span solution to the scale.
Scale
Amount to the scale is approximately 1.3 L.
4. Pour the residual TURB span solution into the cleaning bottle.
Cleaning bottle
Residual TURB span solution
5. Take the detector from the calibration tank, and clean the cell unit using the TURB span
solution in the cleaning bottle.
Clean the cell unit using the TURB span solution
in the cleaning bottle.
CAUTION
Do not spill any solution on the base of cell rotary shaft (V-ring seal part).
6. After cleaning the cell unit with the TURB span solution, dip in the calubration tank.
Allow the solution to adhere by shaking from side to side
for several times.
7. Press the button for [SPAN VALUE] on the TURB. CAL. screen.
The screen changes to the TURB. SPAN screen.
[X.XX]