<8. Maintenance>
8-4
IM 12K01B02-01E
8.2.2
Replacement of Wiper Blade Rubber
Here describes the replacement of wiper blade rubber of measuring cell.
1. Turn ON the maintenance switch.
2. Turn OFF the power switch.
3. Remove the detector from the measuring tank and remove the fixing brackets (2) of the cell
unit.
Press inward.
Press inward.
Bracket
4. Pinch the wiper blade rubber with long-nose pliers etc. and remove it by pulling inward and
downward.
Wiper blade
Wiper blade rubber
Sponge
5. Insert the new wiper blade rubber in the reverse order to step 4. Ensure a correct installing
direction so that the wiping parts or the wiper blade rubber are located in the direction of cell
rotation when viewing from the bottom.
Sponge
Wiping part of wiper
blade rubber
Direction of cell rotation
Sponge
Wiping part of wiper
blade rubber
6. Install the bracket in its original place.
7. Put the detector back in the measuring tank.
8. Turn the power switch ON.
9. Press the [ACTION] button on the MAINTENANCE screen.
Refer to “6.6.1 Action”